Mechanical data

ST 1278 «

TTL

ST 1288 »

1 V

PP

ST 3078 «

TTL

ST 3088 »

1 V

PP

ST 1277 «

TTL

ST 1287 »

1 V

PP

ST 3077 «

TTL

ST 3087 »

1 V

PP

Plunger actuation
Position of plunger at rest

By measured object

Extended

Pneumatic

Retracted

Measuring standard

DIADUR grating on glass; grating period 20 µm

System accuracy

± 1 µm

Position error per signal period

 ± 0.2 µm

Short-range accuracy

typically

0.3 µm

Reference mark

≈ 5 mm below upper stop

Measuring range

12 mm

30 mm

12 mm

30 mm

Working pressure

0.7 bar to 2.5 bars

0.8 bar to 2.5 bars

Radial force

 0.8 N (mechanically permissible)

Fastening

Clamping shank ¬ 8h6

Operating attitude

Any

Vibration

55 Hz to 2 000 Hz

Shock

11 ms

 100 m/s

2

(EN 60 068-2-6)

 1 000 m/s

2

(EN 60 068-2-27)

Operating temperature

10 °C to 40 °C; reference temperature 20 °C

Protection

EN 60 529

IP67/IP64

IP 64

Weight

without cable

40 g

50 g

40 g

50 g

Electrical data

ST 127x
ST 307x

ST 128x
ST 308x

Interface

« TTL

» 1 V

PP

Integrated interpolation*

5-fold

10-fold

Signal period

4 µm

2 µm

20 µm

Edge separation a at
scanning frequency*/traverse speed

2)

100 kHz  72 m/min

1)

50 kHz  60 m/min

25 kHz  30 m/min

 0.48 µs

 0.98 µs

 1.98 µs

 0.23 µs

 0.48 µs

 0.98 µs

Electrical connection*

Cable 1.5 m with D-sub connector (male),

15-pin (integrated interface electronics)

Cable 1.5 m with

• D-sub connector (male), 15-pin

• M23 connector (male), 12-pin

Cable outlet*

Axial or radial

Cable length

 30 m with HEIDENHAIN cable

Voltage supply

5 V DC ± 0.5 V

Current requirement

< 195 mA (without load)

< 55 mA

*

Please select when ordering