HEIDENHAIN Length Gauges User Manual
Page 27

27
Mechanical data
MT 1271 «
TTL
MT 1281 »
1 V
PP
MT 2571 «
TTL
MT 2581 »
1 V
PP
MT 1287 »
1 V
PP
MT 2587 »
1 V
PP
Plunger actuation
Position of plunger at rest
By cable or measured object
Extended
Pneumatic
Retracted
Measuring standard
DIADUR phase grating on Zerodur glass ceramic; grating period 4 µm
System accuracy
± 0.2 µm
Position error per signal period
± 0.02 µm
Short-range accuracy
typically
0.03 µm
0.04 µm
0.03 µm
0.04 µm
Reference mark
≈ 1.7 mm below upper stop
Measuring range
12 mm
25 mm
12 mm
25 mm
Working pressure
–
0.9 bar to 1.4 bars
Radial force
0.8 N (mechanically permissible)
Fastening
Clamping shank ¬ 8h6
Operating attitude
Any; for version without spring and with low gauging force: vertically downward
Vibration
55 Hz to 2 000 Hz
Shock
11 ms
100 m/s
2
(EN 60 068-2-6)
1 000 m/s
2
(EN 60 068-2-27)
Operating temperature
10 °C to 40 °C; reference temperature 20 °C
Protection
EN 60 529
IP50
IP64 (with sealing air)
Weight
without cable
100 g
180 g
110 g
190 g
Electrical data
MT 1271
MT 2571
MT 128x
MT 258x
Interface
« TTL
» 1 V
PP
Integrated interpolation*
5-fold
10-fold
–
Signal period
0.4 µm
0.2 µm
2 µm
Mech. permissible traversing speed
30 m/min
Edge separation a at
scanning frequency*/traverse speed
1)
200 kHz 24 m/min
100 kHz 12 m/min
50 kHz 6 m/min
25 kHz 3 m/min
0.23 µs
0.48 µs
0.98 µs
–
–
0.23 µs
0.48 µs
0.98 µs
–
Electrical connection*
(Interface electronics integrated in
connector)
Cable 1.5 m with D-sub connector (male),
15-pin
Cable 1.5 m with
• D-sub connector (male), 15-pin
• M23 connector (male), 12-pin
Cable length
30 m with HEIDENHAIN cable
Voltage supply
5 V DC ± 0.5 V/< 160 mA (without load)
5 V DC ± 0.25 V/< 130 mA
*
Please select when ordering
1)
At the corresponding cutoff or scanning frequency