Mechanical data

AT 1218

AT 3018

AT 1217

AT 3017

Plunger actuation
Position of plunger at rest

By measured object

Extended

Pneumatic

Retracted

Measuring standard

DIADUR grating on glass; grating period 188.4 µm

System accuracy

± 2 µm

Position error per signal period

 ± 0.7 µm

Measuring range

12 mm

30 mm

12 mm

30 mm

Working pressure

0.7 bar to 1.8 bars

1.1 bars to 1.8 bars

Mech. permissible traversing speed

 80 m/min

 120 m/min

 80 m/min

 120 m/min

Radial force

 0.5 N (mechanically permissible)

Fastening

Clamping shank ¬ 8h6

Operating attitude

Any

Vibration

55 Hz to 2 000 Hz

Shock

11 ms

 100 m/s

2

(EN 60 068-2-6)

 500 m/s

2

(EN 60 068-2-27)

Operating temperature

10 °C to 40 °C; reference temperature 20 °C

Protection

EN 60 529

IP 67

IP 64

1)

IP 67 upon request

IP 64

1)

Weight

without cable

80 g

100 g

80 g

100 g

1)

IP67 with sealing air

Electrical data

EnDat

Interface

EnDat 2.2

Ordering designation

EnDat 22

Resolution

23 nm

368 nm

23 nm

368 nm

Calculation time t

cal

Clock frequency

 5 µs

 8 MHz

Electrical connection

M12 flange socket (male) 8-pin

Cable length

 100 m with HEIDENHAIN cable

Voltage supply

3.6 V to 14 V DC

Power consumption (max.)

3.6 V:  550 mW

14 V:  650 mW

Current consumption (typical)

5 V: 80 mA (without load)