beautypg.com

HEIDENHAIN Length Gauges User Manual

Page 25

background image

25

Specifications

CT 2501

CT 6001

CT 2502

CT 6002

Plunger actuation

By motor

Via coupling with moving machine part

Measuring standard

DIADUR phase grating on Zerodur glass ceramic; grating period 4 µm

System accuracy
At 19 °C to 21 °C

± 0.1 µm,

± 0.03 µm

1)

± 0.1 µm,

± 0.05 µm

1)

± 0.1 µm,

± 0.03 µm

1)

± 0.1 µm,

± 0.05 µm

1)

Position error per signal period

 ± 0.02 µm

Short-range accuracy

typically

0.03 µm

Reference mark

One, ≈ 1.7 mm below upper stop

Measuring range

25 mm

60 mm

25 mm

60 mm

Radial force

 0.5 N (mechanically permissible)

Fastening

Clamping shank

¬ 16h8

Plane surface

Clamping shank

¬ 16h8

Plane surface

Operating attitude

Any required (for preferred operating attitude see Mounting)

Vibration

55 to 2 000 Hz

Shock

11 ms

 100 m/s

2

(EN 60 068-2-6)

 1 000 m/s

2

(EN 60 068-2-27)

Operating temperature

10 °C to 40 °C; reference temperature 20 °C

Protection

EN 60 529

IP50

Weight

without cable

520 g

700 g

480 g

640 g

Electrical data

CT 2501

CT 6001

CT 2502

CT 6002

Interface

» 11 µA

PP

Signal period

2 µm

Measuring velocity

 24 m/min (depending on the subsequent electronics)

 12 m/min with the ND 28x display unit

Electrical connection*

• Cable 1.5 m with D-sub connector (male), 15-pin

• Cable 1.5 m with M23 connector (male), 9 pin

Interface electronics are integrated in connector.

Cable length

 30 m

Voltage supply

5 V DC ± 0.25 V/< 170 mA

5 V DC ± 0.25 V/< 120 mA

Required accessories*

For CT 2501

For CT 6001

Switch box

SG 25 M

SG 60 M

*

Please select when ordering

1)

After linear length-error compensation in the evaluation electronics

2)

Force required to move the plunger or of its weight