Specifications
MT 60 M
MT 101 M
MT 60 K
MT 101 K
Plunger actuation
By motor
Via coupling with moving machine part
Measuring standard
DIADUR grating on silica glass; grating period 10 µm
System accuracy
± 0.5 µm
± 1 µm
± 0.5 µm
± 1 µm
Position error per signal period
± 0.1 µm
Reference mark
≈ 1.7 mm from top
≈ 10 mm from top
≈ 1.7 mm from top
≈ 10 mm from top
Measuring range
60 mm
100 mm
60 mm
100 mm
Radial force
mech. permissible
0.5 N
2 N
0.5 N
2 N
Fastening
Plane surface
Operating attitude
Any
Vertically downward
with SG 101 V
Horizontal with
SG 101 H
Any
Vibration
55 Hz to 2 000 Hz
Shock
11 ms
100 m/s
2
(EN 60 068-2-6)
1 000 m/s
2
(EN 60 068-2-27)
Operating temperature
10 °C to 40 °C; reference temperature 20 °C
Protection
EN 60 529
IP50
Weight
without cable
700 g
1400 g
600 g
1200 g
Electrical data
MT 60 M
MT 101 M
MT 60 K
MT 101 K
Interface
» 11 µA
PP
Signal period
10 µm
Measuring velocity
18 m/min
60 m/min
18 m/min
60 m/min
Electrical connection*
Cable 1.5 m with 15-pin D-sub connector (male) or with 9-pin M23 connector (male)
Cable length
30 m with HEIDENHAIN cable
Voltage supply
5 V DC ± 0.25 V
Current requirement
< 120 mA
< 70 mA
Required accessories*
For MT 60 M
For MT 101 M
Switch box
SG 60 M
Vertical orientation: SG 101 V
Horizontal orientation: SG 101 H
power supply unit
–
Required (see Accessories)
*
Please select when ordering
1)