beautypg.com

HEIDENHAIN ND 281 B v.2 User Manual

Page 60

background image

60

Remote

Operation

over

the

RS-232-C/V.24

Data

Interface

Remote Operation over the RS-232-C/V.24
Data Interface

You can operate the display unit remotely over the RS-232-C/
V.24 data interface. The following commands are available on
the ND 281 B:
Format:
TXXXX Pressed key
AXXXX Output of screen contents
FXXXX Execute function
SXXXX Special function

Sequence of commands

Meaning

T0000

‘0‘ key

T0001

‘1‘ key

T0002

‘2‘ key

T0003

‘3‘ key

T0004

‘4‘ key

T0005

‘5‘ key

T0006

‘6‘ key

T0007

‘7‘ key

T0008

‘8‘ key

T0009

‘9‘ key

T0100

‘CL' key

T0101

‘-‘ key

T0102

‘.‘ key

T0104

‘ENT‘ key

T0105

‘MOD‘ key

T0107

‘1/2‘ (datum) key

Sequence of commands

Meaning

T1000

‘CE+0‘ keys

T1001

‘CE+1‘ keys

T1002

‘CE+2‘ keys

T1003

‘CE+3‘ keys

T1004

‘CE+4‘ keys

T1005

‘CE+5‘ keys

T1006

‘CE+6‘ keys

T1007

‘CE+7‘ keys

T1008

‘CE+8‘ keys

T1009

‘CE+9‘ keys

A0000

Output of model designation

A0100

Output of 14-segment display

A0200

Output of current value

A0301

Output of error text

A0400

Output of software number

A0900

Output of status indicators

F0000

REF function

F0001

Start measurement

1)

F0002

Print

S0000

Counter RESET

S0001

Lock keypad

S0002

Release keypad

1)

Only in the linear measurement mode.