Sorting and Tolerance Checking

In the sorting and tolerance checking mode, the display unit
compares the displayed value with the programmed upper and
lower sorting limits. The sorting and tolerance checking mode is
enabled and disabled with operating parameter P17.

Entering sorting limits

Sorting limits are entered in operating parameters P18 and P19
(see operating parameters).

Sorting signals

The indicators and switching outputs at D-sub connection EXT
(see section on X41) sort the display value into one of three
classes.

Display

Meaning

=

Measured value is within sorting limits

<

Measured value is smaller than lower sorting limit

>

Measured value is greater than upper sorting limit

Operating parameters for sorting and tolerance checking

P17 CLASS

Sorting ON/OFF

P18 L.CLASS.

Lower sorting limit

P19 U.CLASS.

Upper sorting limit

Example: Upper sorting limit = 26.02 mm

Lower sorting limit = 26.00 mm

Part 1

Part 2

Part 3

Sorting and Tolerance Checking

Reject

Accept

Rework