Parameter

Settings / Function

P41 L.COMP.

Linear error compensation

1)

– 99 999.9 < P41 < + 99 999.9 [µm/m]
Default setting: 0

Example: Find input value for P41
Displayed measuring length ............. L

d

= 620.000 mm

Actual length (measured, e.g.
with the VM 101 comparator
system from HEIDENHAIN) .............. L

a

= 619.877 mm

Length difference .............................

DL = L

a

– L

d

= – 123 µm

Compensation factor k (= P41):
k =

DL / L

d

= – 123 µm / 0.62 m ....... k = – 198.4 [µm/m]

P42 BACKLASH

Backlash compensation

1)

Input range (mm):

+9.999 to -9.999

Default setting:

0.000

= no backlash compensation

During a change in direction, the mechanical
play between the encoder and table, the so-
called backlash, can cause error.
Positive backlash: The rotary encoder moves
before the table, therefore the display value
is too high (positive value input).
Negative backlash: The rotary encoder
moves after the table, therefore the display
value is too low (negative value input).

List of Operating Parameters

Parameter

Settings / Function

P43 REF

Reference marks
One reference mark

SINGLE REF.

Distance-coded with 500 • SP
(SP: signal period)

500 SP

Distance-coded with 1000 • SP
(e.g. for HEIDENHAIN LS ...C)

1000 SP

Distance-coded with 2000 • SP

2000 SP

Distance-coded with 5000 • SP

5000 SP

P44 REF

Reference mark evaluation
Evaluate the reference mark

REF. ON

Do not evaluate the reference
mark

REF. OFF

P45 ALARM

Encoder monitoring
No monitoring

ALARM OFF

Monitor the frequency

FREQUENCY

Monitor contamination

CONTAMINAT.

Contamination + frequency

FRQ. + CONT.

P50 RS232

Baud rate

110 / 150 / 300 / 600 / 1200 /

2400 / 4800 / 9600 / 19 200 /

38 400 baud

1)

Only in linear measurement mode.