beautypg.com

Error messages – HEIDENHAIN ND 281 B v.2 User Manual

Page 16

background image

16

Error Messages

Error Messages

Display

Effect/Cause

RS232 FAST

Command for measured value

output followed too quickly by

another.

1)

SIGNAL

Encoder signal is too weak.

The scale may be contaminated.

1)

DSR.MISSING

The connected device has not sent

a DSR signal.

1)

REF. ERR.

The spacing of the reference marks

as defined in P43 is not the actual

spacing.

1)

FORMAT ERR.

Data format, baud rate, etc. do not

match.

1)

FREQUENCY

Input frequency too high for encoder

input. Traversing speed may be

too fast.

1)

MEMORY ERR.

Checksum error: Check the datum,

operating parameters and compen-

sation values for multipoint axis

error compensation. If this error

recurs, contact your service

agency!

1)

These errors are important for the attached device. The
error signal (pin 19) at D-sub connection EXT is active.

Other Error displays
If “OVERFLOW” appears, the measured value is too large or
too small:
ä Set a new datum.

or

ä Traverse back.

If

all sorting signals light up, the upper sorting limit is smaller

than the lower limit:
ä Change operating parameters P18 and/or P19.

To clear error messages:

Once you have removed the cause of the error:
ä Clear the error message with the CL key.

Display

Effect/Cause

REC. ERROR

Error during reception of param-

eter and compensation-value lists.