List of operating parameters – HEIDENHAIN ND 281 B v.2 User Manual
Page 24

24
1)
Only in linear measurement mode.
2)
Only in angular measurement mode.
Parameter
Settings / Function
P09 ANGLE
Angle display
2)
+/- 180°
+/- 180 DEG.
+/-
¥
ENDLESS
360°
360 DEG
.
P11 SCL
Scaling factor
1)
Scaling factor off
SCALING OFF
Scaling factor on
SCALNG. ON
P12 SCL
Scaling factor
1)
Enter a numerical value
0.100000 < P12 < 9.999999
Default setting: 1.000000
P17 CLASS
Sorting and tolerance checking
Sorting and tol. checking ON
CLASS ON
Sorting and tol. checking OFF
CLASS OFF
P18 L.CLASS
Lower limit for sorting
P19 U.CLASS
Upper limit for sorting
P21 SERIES
Display for series of measurements
1)
OFF MIN MAX ACTL DIFF
Parameter
Settings / Function
P00 CODE
Enter a code number:
9 51 48:
To change the protected
operating parameters
41 52 63: To select the linear or angular
measurement mode
10 52 96: Multipoint axis error
compensation
24 65 84: To lock the keyboard
66 55 44: To show the software version
24 65 82: Distance-to-go display
48 61 53: Input and output of parameter
and compensation-value lists
P01
Units of measurement
1)
Display in millimeters
MM
Display in inches
INCH
P02 X1/X2
Select an encoder input
Signals at X1
11 µAPP
Signals at X2
1 VPP
P08 DISPL.
Display mode
2)
Decimal degrees
DEC. DEGREE
Deg., minutes, seconds
DEG.MIN.SEC.
List of Operating Parameters
List of Operating Parameters