Parameter

Settings/ Function

P37 STEP

Counting mode

2)

0-1-2-3-4-5-6-7-8-9

COUNT 0-1

0-2-4-6-8

COUNT 0-2

0-5

COUNT 0-5

P38 DEC.

Decimal places

3)

1 / 2 / 3 / 4 / 5 / 6
(up to 8 for inch display)

P40 COMP.

Select encoder compensation
No compensation

COMP. OFF

Multipoint: up to 64 compensation points for
linear encoders, up to 72 for angle encoders
(fixed spacing of 5 deg.)

COMP. MULTI

Linear compensation

COMP. LIN

1)

Parameter

Settings/ Function

P23 DISPL.

Display stop for measured value output
Concurrent display
, no display freeze;
the display value is the current actual
value

DISPL. ACTL.

Frozen display; hold display until next
measured value output

DISPL. HOLD

Frozen/concurrent display; freeze display
as long as Pulse/Contact for measured
value output is present

DISPL. STOP

P30 DIR

Counting direction
Positive counting direction with
positive direction of traverse

DIRECT. POS

Negative counting direction with
positive direction of traverse

DIRECT. NEG

P31 S. PER.

Signal period

1)

of encoder

0.000 000 01 < P31 < 99 999.9999
Default setting: 10 µm

P33 COUNT

Counting mode

1)

0-1-2-3-4-5-6-7-8-9

COUNT 0-1

0-2-4-6-8

COUNT 0-2

0-5

COUNT 0-5

P36 SP/R

Signal periods per revolution

2)

1 < P36 < 999 999
Default setting: 36 000

List of Operating Parameters

1)

Only in linear measurement mode.

2)

Only in angular measurement mode.

3)

Depends on signal period (P31) and unit of measure (P01)