beautypg.com

Distance-to-go mode – HEIDENHAIN ND 281 B v.2 User Manual

Page 45

background image

45

Distance-to-Go Mode

Distance-to-Go Mode

1)

Normally, the display shows the actual position of the en-
coder. However, it is often more helpful to display the re-
maining distance to an entered nominal position—especially
when you are using the display unit for machine tools and au-
tomation purposes. You can then position simply by traversing
to display value zero.

You can access the distance-to-go display by entering the
code number 24 65 82.

Function of switching outputs A1 and A2

In the distance-to-go mode, switching outputs A1 (pin 15) and
A2 (pin 16) have a different function: they are symmetrical to
the display value zero. For example, if a switching point of
10 mm is entered in P62, output A1 switches at both +10 mm
and –10 mm. The figure below shows output signal A1 when
approaching zero from the negative direction.

Time curve of a signal for switching limit (A1) = 10 mm,
t

d1

£ 30 ms, t

d2

£ 180 ms

Display

Meaning

DELTA ON

Distance-to-go display active

DELTA OFF

Distance-to-go display not active

“Traversing to zero” with distance-to-go display

ä Select datum point 2.

ä Enter the nominal position.
➤ Move the axis until the display value is zero.

1)

Only in linear measurement mode

t

d2

t

d1