5setup procedures, 1 conductivity meter – AEC FSS Series 1.5- 10 HP Fountain Solution Systems User Manual
Page 43

Page 42
FSS Series Fountain Solution Systems
IAE-600
5
Setup Procedures
5-1 Conductivity Meter
GLI PRO-SERIES
1) Press the menu key.
2) Scroll down to configure and press enter.
3) Scroll down to set C (Celsius) or F (Farenheight) and press
enter.
4) Scroll up to change C (Celsius) to F (Farenheight) and press
enter.
5) Scroll down to sensor and press enter.
6) Scroll down to display format and press enter.
7) The screen should read display format.
8) Press the arrow up or down until screen reads <20.00 mS/CM>
and press enter.
9) Scroll down to temp element and press enter.
10) Scroll down to Set “T” factor? and press enter.
11) Look at the label on the sensor to get the resistance value. (eg:
T=1000.5)
12) Enter the value found on the sensor and press enter.
13) Scroll down to exit then press enter. (This will take you back
to the sensor screen)
14) Scroll down to exit again and press enter. (This will take you
back to the configure screen)
15) Scroll down to set output and press enter.
16) Scroll down to 4 mA value and press enter.
17) Change to 0.00 mS/CM and press enter.
18) Scroll down to 20.00 mA and press enter.
19) Set the 20 mA value to 5.00 mS/CM and press enter.
20) Scroll down to exit then press enter. (This will take you back
to the configure screen)
21) Scroll down to exit and press enter. (This will take you back
to the main menu screen)
22) Scroll down to calibrate then press enter.
23) Sensor should appear then press enter.
24) Scroll down to zero then press enter.