Sloan 111 Crown/Crown II Exposed Flushometer User Manual
Page 5

5
1. PROBLEM: Valve does not function.
CAUSE: (a.) Control Stop or Main Valve is Closed.
(b.) Handle Assembly is worn.
(c.) Relief Valve is worn and sticking in UP position.
SOLUTION: (a.) Open Control Stop or Main Valve.
(b.) Install Sloan Handle Repair Kit (CR-1007) or replace
Handle.
(c.) Replace Piston.
2. PROBLEM: Insufficient volume of water to adequately
siphon fixture.
CAUSE: (a.) Control Stop not open enough.
(b.) Urinal Piston Parts inside a Closet Valve.
(c.) Low Consumption Valve installed on a non-Low
Consumption fixture.
(d.) Inadequate volume or pressure at supply.
SOLUTION: (a.) Adjust Control Stop for desired delivery of water.
(b.) Replace Piston Parts with proper Closet Piston.
(c.) Replace with proper Flushometer. Crown II and Crown
Flushometers are not available with flush volume
higher than 1.6 gpf/6.0 Lpf.
(d.) Increase water pressure or supply (flow) to valve.
Consult Factory for assistance.
IMPORTANT - LAWS AND REGULATIONS
PROHIBIT THE USE OF HIGHER
FLUSHING VOUMES THAN LISTED ON
FIXTURE OR FLUSHOMETER.
3. PROBLEM: Length of flush too short (Short Flushing) or
Valve closes off immediately.
CAUSE: (a.) Piston Assembly is not hand-tight.
(b.) Enlarged Bypass orifice from corrosion or damage.
(c.) Urinal Piston in Closet Flushometer.
(d.) Low Consumption Valve installed on a non-Low
Consumption fixture.
(e.) Handle Assembly worn.
SOLUTION: (a.) Screw the Piston Assembly hand-tight.
(b.) Install NEW Inside Parts Kit to correct problem and
update flushometer.
(c.) Replace Piston with proper Closet Piston.
(d.) Replace with proper flushometer. Crown II and Crown
flushometers are not available with flush volume
higher than 1.6 gpf/6.0 Lpf.
(e.) Install Handle Repair Kit (CR-1007).
4. PROBLEM: Length of flush too long (Long Flushing) or
fails to close off.
CAUSE: (a.) Piston is not seating properly or Bypass orifice is
clogged because of foreign material, or Bypass orifice
is clogged by an invisible gelatinous film from “over-
treated” water.
(b.) Line pressure has dropped and is not sufficient to
force Relief Valve to seat.
(c.) Main Seat is fouled with debris or is worn.
SOLUTION: (a.) Disassemble the working parts and wash thoroughly.
NOTE: Size of the orifice in the Bypass is of utmost
importance for the proper metering of water into
the upper chamber of the valve. Do not enlarge or
damage this orifice. Replace Piston if cleaning does
not correct problem.
(b.) Shut off all control stops until pressure has been
restored, then open them again.
(c.) Clean or replace Main Seat.
5. PROBLEM: Leaking at Handle Assembly.
CAUSE: (a.) The B-85 Seal may have deteriorated.
(b.) Handle Gasket has been omitted.
(c.) Valve Handle Bushing is worn.
(d.) Handle Trim Ring is reversed preventing gasket from
sealing (Crown II only).
SOLUTION: (a.) Install new B-85 Seal.
NOTE: The B-85 Seal will easily slide onto the
Bushing if it is wet.
(b.) Install A-31 Handle Gasket or Sloan Handle Repair Kit
(CR-1007).
(c.) Install Sloan Handle Repair Kit (CR-1007).
(d.) Flip Handle Trim Ring (CR-113) and reinstall Handle
Assembly.
Refer to the Repair Parts and Maintenance Guide for additional
Troubleshooting and Repair Part information.
If further assistance is required,
please contact Sloan Technical Support at:
1-888-SLOAN-14 (1-888-756-2614)
TROUBLESHOOTING GUIDE