background image

NOTE: P = Pass test.

S = Skip, test not performed.
F = Fail test.
T = Terminate test if load or leakage pretest fail.
A = Aborted test by user.

$ = Alphanumeric character.

xx . xx = Measured result value.

DATA l/O

A 9 pin ‘D’ type connector is located at the right hand end of the keypad.

The data output uses RS232 levels using ±5VDC The pin connections are as follows:

Pin 517 Ground, Earth

Pin 3 Data out

Pin 2 Busy/Data in

Pin 1,4,6,8 NO connection

Pin 9 +5V (0.5A max)

The set-up data for computer/printer is as follows:

Baud rate 9600 for computer l/0,1200 for printer output.

Start bits 1

Stop bits 2

Data bits 8

No parity

For a bar code reader the data set-up is as follows:

Baud rate 9600

Stop bits 2

Data bits 8

Intercharacter delay 20mS

HP BAR CODE READER SET- UP

At the prompt ‘PRESS # FOR NEW TEST’ plug the barcode reader into the serial port
on the KEYPAD. Press the ‘W’ key on the keyboard.
The instrument will now automatically send the required codes for bar code reader set-
up and the bar code reader will now be configured.
Test the barcode reader on a suitable barcode for verification.

PAT 2000i

Portable Appliance Tester

22

5 4 3 2 1

9 8 7 6

Serial Data Connector View

Looking at Socket

PAT 2000i Booklet 7/3/03 4:26 pm Page 23