9.Technical Specifications

73

9. Technical Specifications

Unless otherwise specified, the technical specifications applied are for TDS series
only, and Probes attenuation set as 10X. Only if the oscilloscope fulfills the following
two conditions at first, these specification standards can be reached.

 This instrument should run for at least 30 minutes continuously under the

specified operating temperature.

 If change of the operating temperature is up to or exceeds 5℃, do a

"Self-calibration" procedure (see "How to Implement Self-calibration" on P14).

All specification standards can be fulfilled, except one(s) marked with the word
"Typical".

Performance Characteristics

Instruction

Bandwidth

TDS7104
TDS8104

100MHz

TDS8204

200MHz

Channel

4

Acquisition

Mode

Normal, Peak detect, Averaging

Waveform Capture
rate

50,000

Max Sample rate
(real time)

TDS7104

Four CH

500MS/s

Dual CH

1GS/s*

Single CH

1GS/s

TDS8104
TDS8204

Four CH

1GS/s

Dual CH

2GS/s*

Single CH

2GS/s

Input

Input coupling

DC, AC , Ground

Input impedance

1M

Ω±2%, in parallel with 10pF±5pF

50Ω±1%

Probe attenuation

factor

1X, 10X, 100X, 1000X

Max. input voltage

1MΩ input impedance:
400V (PK-PK) (DC + AC PK-PK)
50Ωinput impedance:
5V (PK-PK) (DC + AC PK-PK)

Bandwidth limit

20MHz, full bandwidth

Channel

–channel

isolation

50Hz: 100 : 1
10MHz: 40 : 1