5.Advanced User Guidebook

32

How to Set the Trigger System

Trigger determines when TDS starts to acquire data and display waveform. Once
trigger is set correctly, it can convert the unstable display to meaningful
waveform.
When TDS starts to acquire data, it will collect enough data to draw waveform
on left of trigger point. TDS continues to acquire data while waiting for trigger
condition to occur. Once it detects a trigger it will acquire enough data
continuously to draw the waveform on right of trigger point.
Trigger control area consists of 1 knob and 3 menu keys.
TRIG LEVEL: The knob that set the trigger level; press the knob and the level

will be cleaned to Zero.

50%: The instant execute button setting the trigger level to the vertical midpoint

between the peaks of the trigger signal.

Force: Force to create a trigger signal and the function is mainly used in

"Normal" and "Single" mode.

Trigger Menu: The button that activates the trigger control menu.

Trigger Control

Trigger: Use a trigger level to capture stable waveforms in 4 channels

simultaneously.

Trigger has four modes: edge trigger, video trigger, slope trigger and pulse trigger.
Edge Trigger: It occurs when the trigger input passes through a specified voltage

level with the specified slope.

Video Trigger: Trigger on fields or lines for standard video signal.

Slope Trigger: The oscilloscope begins to trigger according to the signal rising or

falling speed.

Pulse Trigger: Find pulses with certain widths.

The four trigger modes in Trigger are described respectively as follows:

1. Edge Trigger

An edge trigger occurs on trigger threshold value of input signal. Select Edge trigger
mode to trigger on rising edge or falling edge.

The description of the Edge Trigger Menu is as follows:

Menu

Settings Instruction

Trigger Mode

Edge

Set vertical channel trigger type as edge trigger.

Source




AC

Channel 1 as trigger signal.
Channel 2 as trigger signal.
Channel 3 as trigger signal.
Channel 4 as trigger signal.
AC power line as trigger signal.