beautypg.com

The automatic measurement of time parameters – OWON TDS Series User Manual

Page 56

background image

5.Advanced User Guidebook

51

entire waveform.

Vamp: Voltage between Vtop and Vbase of a waveform.
Vtop: Voltage of the waveform's flat top, useful for square/pulse waveforms.
Vbase: Voltage of the waveform's flat base, useful for square/pulse waveforms.
Overshoot: Defined as (Max-Vtop)/Vamp, useful for square and pulse

waveforms.

Preshoot: Defined as (Min-Vbase)/Vamp, useful for square and pulse waveforms.
Mean: The arithmetic mean over the entire waveform.
RMS: The true Root Mean Square voltage over the entire waveform.

The automatic measurement of time parameters

The TDS series oscilloscopes provide time parameters auto-measurements
include Frequency, Period, Rise Time, Fall Time, +Width, -Width, +Duty, -Duty,
Delay

, Delay

, Delay

, Delay

.

Figure 5-9 shows a pulse with some of the time measurement points.

Rise Time

Fall Time

+Width

-Width

Figure 5-9

Rise Time: Time that the leading edge of the first pulse in the waveform takes to

rise from 10% to 90% of its amplitude.

Fall Time: Time that the falling edge of the first pulse in the waveform takes to

fall from 90% to 10% of its amplitude.

+Width: The width of the first positive pulse in 50% amplitude points.

-Width: The width of the first negative pulse in the 50% amplitude points.

Delay

and Delay

: The delay between the two channels at the

rising edge.

Delay

and Delay

: The delay between the two channels at the

falling edge.

+Duty: +Duty Cycle, defined as +Width/Period.

-Duty:-Duty Cycle, defined as -Width/Period.