Created: 21.09.2006
Changed: 29.02.2008
Printed: 02.03.2009

page 38 of 59

voltage supply PerPur:

120VAC, 10%, 1 Phase

fuse protection:

15A, surge proof

wires:

L1, N, PE

cable cross section:

1.5 mm² (16 AWG)


The remote fault indicator can be connected by means of a 2 pole plug with an optical or
acoustic transmitter or with a building services management system.

5.4.3

Pressure Booster


The voltage supply is established via a wall socket and the connecting cable at the DE 3-5,
which has a 3 pole shockproof plug.

supply voltage DE 3-5:

120VAC, 10%, 1 Phase

fuse protection:

10A, surge proof

wires:

L1, N, PE

cable cross section:

1.5 mm² (16 AWG)