Pipe – Nortec DRAABE Pure Water System User Manual
Page 31
Created: 21.09.2006
Changed: 29.02.2008
Printed: 02.03.2009
page 31 of 59
Pipe
When mounting the alternating current connection be sure the wall socket is mounted ap-
proximately 25 cm (9.84 inches) to the left of the wall bracket and at the height of the pure
water system.
Th
reads: 3/8”, female
Figure 14: Mounting the PerPur Console
Together with the console
three fittings are delivered
which can be used to con-
nect this unit by a PVC or
CPVC pipe (16mm outer di-
ameter).
Upward clearance of 50cm (20 inches) across the entire width of the console should be
provided The clearance to the right should be 20cm (8 inches). The distance from the lower
edge to the floor should be 1.2m (47 inches).
CAUTION:
No parts which conduct electricity may be mounted under the console because leakage
could cause damages.
5.3.4
Pure Water Tank (wall model)
-
Mounting on a wall that can bear loads is required (weight ~ 100kg).
-
If at all possible, the device should be installed in a room to which only a limited cate-
gory of persons is granted access.
-
The location must be selected so that direct sunlight and heat radiation onto the device
is impossible at all times.
516 mm
(20 inches)
185 mm (7.3 inches)
258 mm
(11 inches)