Operating guidelines – Measurement Computing SWITCH User Manual
Page 25

Switch & Sense 8 User's Guide Template
Functional Details
25
Operating guidelines
Follow these guidelines to help make your Switch & Sense 8 run as trouble-free as possible.
Run the Switch & Sense 8 software (ICalUSB.exe) after you connect each Switch & Sense 8 for the first
time. ICalUSB.exe generates a configuration file used to map the specified device number to the physical
device.
Do not disconnect any device from the USB bus while the host is communicating with the Switch & Sense
8.
If the
STATUS
LED to the left of the USB cable goes out, your computer cannot communicate with the
device. To re-enable the device, disconnect the USB cable from the computer, and then reconnect it. This
action resets the
STATUS
LED to on and restores communication.
If using the relays to control inductive loads, place a diode across the control terminals to suppress the
kickback voltage. If the diode is not present, the kickback voltage could reset the on-board processor. The
only way to recover from a reset is to disconnect the USB cable, then reconnect it.