beautypg.com

Measurement Computing eZ-Balance rev.9.0 User Manual

Page 26

background image

3-8 Windows, Buttons, and Menus

969391

eZ-Balance

Edit Menu > Configuration Window >

Input Channels Tab

The Balance Configuration window includes an Input Channels tab (following figure). Use this tab to
configure each input signal. Set the values for each attribute on a channel by channel basis. Note that the
following figure does not show all attributes. To view additional attributes you would use the scroll bar
[below channel 8] and scroll to the right. A table of channel attributes follows.

Balance Configuration Input Channels Tab

Configuration Window . . . Input Channels Tab

Attribute

Comments

Type

This is the type of instrument that will be connected to the selected channel.

Instrument Types include: Accelerometer, Velocity, X probe, Y probe, Z probe, and
Tach. Make your selection on the pull down menu. If no instrument will be
connected to the channel select “unused.”

You must have 1 Tach Channel and 1 or more Response Channels.

For IOtech 650 units, CH5 should never be configured as an accelerometer channel

since it has no IEPE capability.

Name

Enter a name, or label, for the selected channel.

Gauge Min
Gauge Max

Enter the instrument Range: Minimum and Maximum in engineering units.

Units

Default engineering units are based on the Instrument Type selected. To change the
units, double-click on the cell and type in the desired unit label.

In regard to selecting an IOtech 650 unit channel Type [preceding table], channel 5
should never be configured for accelerometer input. This is because channel 5 has
no IEPE capability.

This manual is related to the following products: