11.
Apply an acceptable “Second Point”
calibrator voltage for the range, then
hit
the Finish button on your screen).
WaveCal is now ready for the “first
point” calibration of the next range.
12.
Continue this method for the
remaining ranges and for each
channel. Remember, if Concurrent is
selected (on the Workbook
Calibration Tool window, the
Channel box will show 1-8 and the
eight channels will be calibrated
simultaneously.
Channel 1, Range –2.5 to 2.5 V
Scheduled for
Second Point Calibration Using +2.4 V
WaveCal allows you to select
individual channels and ranges for
calibration. For example:
If you have a channel with
questionable values for one range,
but all other channels and ranges
appear fine, you can:
a)
Use “non-concurrent” (no
check-box in Concurrent)
b)
Select the applicable channel.
c)
Select the range that is in
question.
d)
Calibrate for First Point.
e)
Calibrate for Second Point.
Channel 1, Range –1.0 to 1.0 V
Selecting a Specific Channel and Range for Calibration