Section 2

Getting Started

2.7

command/status and serial data addresses w hen using the dual
primary addressing mode.

Figure 2-5 Pictorial View of the Dual Primary Addressing

Mode

Port 2

Control /

Status

Port 1
Control /
Status

Port 3
Control /
Status

Port 4
Control /
Status

Port 2
Data

Port 1
Data

Port 3
Data

Port 4
Data

Port Select Command

IEEE 488

Even Address (08)
(command )

Odd Address (09)
(data )

2.3.2 IEEE 488 Address Selection

The IEEE 488 bus address is set by SW1-1 through SW1-5 (located

on rear panel). The address is read only at pow er on and can be set from 0
through 30 when using the secondary addressing mode, or 0 to 28 using the
dual primary addressing mode. The address is selected by simple binary
weighting with SW1-1 being the least significant bit and SW 1-5 the most
significant bit (see Figure 2-6). SW1 switch positions for IEEE addresses 0
through 30 are listed in Table 2-1. The factory default is address 8.

NOTE: If address 30 / 31 is selected, when using dual primary addressing
mode, it defaults to address 28 / 29 because the IEEE 488 standard has
reserved address 31. Similarly if address 31 is selected, when using the
secondary addressing mode, it defaults to address 30.