beautypg.com

Pdq30 specifications, General, Analog inputs – Measurement Computing Personal Daq/3000 Series User Manual

Page 77

background image

Note

: Specifications are subject to change without notice.

937492

Specifications 7-11

PDQ30 Specifications

General

Operating Temperature:

-30° to +70°C

Storage Temperature:

-40° to +80°C

Power:

Supplied by DaqBoard/3000 Series or Personal Daq/3000 Series; 400 mW (max)

Relative Humidity:

0 to 95%, non-condensing

Vibration:

MIL STD 810E, category 1 and 10

Communications Connector:

25 pin DSUB

Signal I/O Connector:

Six removable screw-terminal blocks (12 connections each)

Dimensions

: 269mm W x 92mm D x 45 mm H: (10.6” x 3.6” x 1.6”)

Weight

: 400g (0.88 lbs)

Analog Inputs

Channels

: 48 single-ended inputs; or 24 channels differential inputs

Voltage Measurement Speed

: 1µs per channel

Ranges

: ±10V, ±5V, ±2V, ±1V, ±500mV, ±200mV, ±100mV, universal thermocouple.

Software or sequencer selectable on a per-channel basis

Total Harmonic Distortion

: -80dB typical for ±10V range, 1 kHz fundamental

Signal to Noise and Distortion

: 72dB typical for ±10V range, 1 kHz fundamental

Input Impedance

: 10M Ohm (single-ended); 20M Ohm (differential)

Bias Current

: 40 pA typical (0 to 35°C)

Over-Voltage Protection

: ±30V

Crosstalk

: -75 dB DC to 60 Hz; -65 dB @ 10 kHz, typical

Voltage

Range*

Accuracy

±(% Reading + % Range)

23ºC ± 10°C, 1 year

Temperature Coefficient

± (ppm of Reading + ppm Range)/ºC

-30ºC to 13°C and 33ºC to 70ºC

Noise**

(cts RMS)

-10V to 10V

0.031% + 0.008%

14 + 8

2.0

-5V to 5V

0.031% + 0.009%

14 + 9

3.0

-2V to 2V

0.031% + 0.010%

14 + 10

2.0

-1V to 1V

0.031% + 0.02%

14 + 12

3.5

-500 mV to 500 mV

0.031% + 0.04%

14 + 18

5.5

-200 mV to 200 mV

0.036% + 0.05%

14 + 12

8.0

-100 mV to 100 mV

0.042% + 0.10%

14 + 18

14.0

* Specifications assume differential input single channel scan, 1-MHz scan rate, unfiltered, CMV=0.0V,

30 minute warm-up; exclusive of noise, range -FS to +FS.

** Noise reflects 10,000 samples at 1-MHz, typical, differential short

TC Types and Accuracy

1

TC

Temp Range (°C)

Accuracy (±°C)

Noise (±°C)

J

-200 to

+760

1.7

0.2

K

-200 to

+1200

1.8

0.2

T

-200 to

+400

1.8

0.2

E

-270 to

+650

1.7

0.2

R

-50 to

+1768

4.8

1.5

S

-50 to

+1768

4.7

1.5

N

-270 to

+1300

2.7

0.3

B

+300 to +1400

3.0

1.0

1

Assumes 16384 oversampling applied, CMV = 0.0V, 60 minute warm-up, still environment, and 25°C

ambient temperature; excludes thermocouple error; TC

IN

= 0°C for all types except B (1000°C), TR-2 for

External Power.