7-4 Specifications

937492

Note

: Specifications are subject to change without notice.

Voltage

Range*

Accuracy

±(% Reading + % Range)

23ºC ± 10°C, 1 year

Temperature Coefficient

± (ppm of Reading + ppm Range)/ºC

-30ºC to 13°C and 33ºC to 70ºC

Noise**

(cts RMS)

-10V to 10V

0.031% + 0.008%

14 + 8

2.0

-5V to 5V

0.031% + 0.009%

14 + 9

3.0

-2V to 2V

0.031% + 0.010%

14 + 10

2.0

-1V to 1V

0.031% + 0.02%

14 + 12

3.5

-500 mV to 500 mV

0.031% + 0.04%

14 + 18

5.5

-200 mV to 200 mV

0.036% + 0.05%

14 + 12

8.0

-100 mV to 100 mV

0.042% + 0.10%

14 + 18

14.0

* Specifications assume differential input single-channel scan, 1-MHz scan rate, unfiltered,

CMV=0.0V, 30 minute warm-up, exclusive of noise, range is -FS to +FS.

** Noise reflects 10,000 samples at 1-MHz, typical, differential short.

TC Types and Accuracy

1

TC Type

Temperature

Range (°C)

Accuracy

(±°C)

Noise,

Typical

(±°C)

J

-200 to +760

1.7

0.2

K

-200 to +1200

1.8

0.2

T

-200 to +400

1.8

0.2

E

-270 to +650

1.7

0.2

R

-50 to +1768

4.8

1.5

S

-50 to +1768

4.7

1.5

N

-270 to +1300

2.7

0.3

B

+300 to +1400

3.0

1.0

1

Assumes 16384 oversampling applied, CMV = 0.0V, 60 minute warm-up, still environment, and

25°C ambient temperature; excludes thermocouple error; TC

IN

= 0°C for all types except B

(1000°C), TR-2 for External Power.

Oversampling Effect on Measurement Noise

0.1

1

10

-300

-200

-100

0

100

200

300

400

Measured Temperature (ºC)

P

-P

N

o

ise

in M

e

as

ur

em

ent

C)

Type T Thermocouple (Lines represent theoretical noise)

16384

8291

4096

256

512

1024

2048