Triggering – Measurement Computing Personal Daq/3000 Series User Manual
Page 73

Note
: Specifications are subject to change without notice.
937492
Specifications 7-7
Triggering
Trigger Sources
: 7, individually selectable for starting and stopping an acquisition. Stop acquisition can occur on
a different channel than start acquisition; stop acquisition can be triggered via modes 2, 4, 5, or 6 described
below.
1. Single-Channel Analog Hardware Trigger
: Any analog input channel can be software programmed as
the analog trigger channel, including any of the analog expansion channels.
Input Signal Range
: -10 to +10V max
Trigger Level
: Programmable; 12-bit resolution
Hysteresis
: Programmable; 12-bit resolution
Latency
: 350 ns typical, 1.3 µs max
Accuracy
: ±0.5% of reading, ±2 mV offset
Noise
: 2 mV RMS
2. Single-Channel Analog Software Trigger
: Any analog input channel, including any of the analog
expansion channels, can be selected as the software trigger channel. If the trigger channel involves a
calculation, such as temperature, then the driver automatically compensates for the delay required to
obtain the reading, resulting in a maximum latency of one scan period.
Input Signal Range
: Anywhere within the range of the selected trigger channel
Trigger Level
: Programmable; 16-bit resolution, including “window triggering”
Latency
: One scan period max
3. Single-Channel Digital Trigger
: A separate digital input is provided for digital triggering.
Input Signal Range
: -15V to +15V
Trigger Level
: TTL
Minimum Pulse Width
: 50 ns high; 50 ns low
Latency
: 100 ns typical, 1.1 µs max
4.
Digital Pattern Trigger: 8 or 16-bit pattern triggering on any of the digital input ports. Programmable for
trigger on equal, above, below, or within/outside of a window. Individual bits can be masked for “don’t
care” condition.
Latency
: One scan period max
5.
Counter/Totalizer Trigger: Counter/totalizer inputs can trigger an acquisition. User can select to trigger
on a frequency or on total counts that are equal, above, below, or within/outside of a window.
Latency
: One scan period max
6. Software Trigger
: Trigger can be initiated under program control.
7. Multi-Channel Triggering
: Up to 16 channels can be used to generate a trigger condition for any
combination of analog, digital, or counter inputs. Multiple channels can either be combined in a logical
“or” or “and” condition, with hysteresis programmable per channel. Maximum latency in this mode is
one scan period.