Triggering

Trigger Sources

: 7, individually selectable for starting and stopping an acquisition. Stop acquisition can occur on

a different channel than start acquisition; stop acquisition can be triggered via modes 2, 4, 5, or 6 described

below.

1. Single-Channel Analog Hardware Trigger

: Any analog input channel can be software programmed as

the analog trigger channel, including any of the analog expansion channels.
Input Signal Range

: -10 to +10V max

Trigger Level

: Programmable; 12-bit resolution

Hysteresis

: Programmable; 12-bit resolution

Latency

: 350 ns typical, 1.3 µs max

Accuracy

: ±0.5% of reading, ±2 mV offset

Noise

: 2 mV RMS

2. Single-Channel Analog Software Trigger

: Any analog input channel, including any of the analog

expansion channels, can be selected as the software trigger channel. If the trigger channel involves a

calculation, such as temperature, then the driver automatically compensates for the delay required to

obtain the reading, resulting in a maximum latency of one scan period.
Input Signal Range

: Anywhere within the range of the selected trigger channel

Trigger Level

: Programmable; 16-bit resolution, including “window triggering”

Latency

: One scan period max


3. Single-Channel Digital Trigger

: A separate digital input is provided for digital triggering.

Input Signal Range

: -15V to +15V

Trigger Level

: TTL

Minimum Pulse Width

: 50 ns high; 50 ns low

Latency

: 100 ns typical, 1.1 µs max

4.

Digital Pattern Trigger: 8 or 16-bit pattern triggering on any of the digital input ports. Programmable for

trigger on equal, above, below, or within/outside of a window. Individual bits can be masked for “don’t

care” condition.
Latency

: One scan period max

5.

Counter/Totalizer Trigger: Counter/totalizer inputs can trigger an acquisition. User can select to trigger

on a frequency or on total counts that are equal, above, below, or within/outside of a window.
Latency

: One scan period max

6. Software Trigger

: Trigger can be initiated under program control.


7. Multi-Channel Triggering

: Up to 16 channels can be used to generate a trigger condition for any

combination of analog, digital, or counter inputs. Multiple channels can either be combined in a logical

“or” or “and” condition, with hysteresis programmable per channel. Maximum latency in this mode is

one scan period.