Measurement Computing DaqBoard 3000USB Series User Manual
Page 97

Note: Specifications are subject to change without notice.
928391
Specifications 7-3
Worst Case Temperature Measuerment Error vs. PDaq3000 Ambient Temperature
with Thermocouple at 0ºC (Excludes Thermocouple Error)
AutoZero Disabled
0
1
2
3
4
5
6
7
8
-30
-20
-10
0
10
20
30
40
50
60
70
Ambient Temperature (ºC)
Te
m
pe
ra
ture
Me
a
s
ure
m
e
nt
E
rror (
ºC)
f
or Ty
pe
s
T, J
, K, E
,
R, S
J
K
T
E
Voltage Measurement Noise vs. Oversampling
0.01
0.10
1.00
10.00
100.00
1
10
100
1000
10000
100000
Oversampling Rate
Nois
e
(
RMS
Count
s
), Ty
pic
a
l
100mV
10V
2V
1V
500mV
200mV
Range
Note: does not include noise due to quantizing of returned data
See also other documents in the category Measurement Computing Hardware:
- ACC-300 (7 pages)
- AI-EXP32 (20 pages)
- AI-EXP48 (19 pages)
- BTH-1208LS (30 pages)
- 6K-ERB08 (32 pages)
- BTH-1208LS Quick Start (4 pages)
- 6K-SSR-RACK08 (33 pages)
- BTH-1208LS-OEM (27 pages)
- CB-COM-Digital (68 pages)
- CB-7018 (68 pages)
- CB-7000 Utilities (44 pages)
- CB-7080D (74 pages)
- CB-COM-7033 (44 pages)
- CB-COM-7017 (72 pages)
- CB-COM-7024 (76 pages)
- CB-NAP-7000P (36 pages)
- CIO-DAC02/16 (16 pages)
- CIO-DAC02 (18 pages)
- CB-NAP-7000D (56 pages)
- CIO-DAC16-I (16 pages)
- CIO-DAC16/16 (20 pages)
- CIO-DAS08 (21 pages)
- CIO-DAC16 (20 pages)
- CIO-DAS08/JR (16 pages)
- CIO-DAS08/JR/16 (14 pages)
- CIO-DAS08/JR-AO (16 pages)
- CIO-DAS08-AOM (32 pages)
- CIO-DAS08-PGM (28 pages)
- CIO-DAS16/330 (34 pages)
- CIO-DAS48-I (17 pages)
- CIO-DAS16/M1 (38 pages)
- CIO-DAS48-PGA (18 pages)
- CIO-DAS800 (20 pages)
- CIO-DAS802/16 (22 pages)
- CIO-DAS6402/16 (40 pages)
- CIO-DAS-TEMP (20 pages)
- CIO-DDA06/16 (18 pages)
- CIO-DDA06/JR (17 pages)
- CIO-DIO24H (20 pages)
- CIO-DIO24/CTR3 (21 pages)
- CIO-DI192 (24 pages)
- CIO-DDA06 (21 pages)
- CIO-DIO48 (19 pages)
- CIO-DO192H (16 pages)
- CIO-DIO192 (20 pages)