beautypg.com

Analog outputs, Digital i/o – Measurement Computing DaqBoard 3000USB Series User Manual

Page 101

background image

Note: Specifications are subject to change without notice.

928391

Specifications 7-7

Analog Outputs

/3001USB and /3031USB models only

Analog output channels are updated synchronously relative to scanned inputs, and clocked from either an
internal onboard clock, or an external clock source. Analog outputs can also be updated asynchronously,
independent of any other scanning in the system. Streaming from disk or memory is supported, allowing
continuous waveform outputs (limited only by available PC system resources).

Channels:

4 DAC channels (DAC0, DAC1, DAC2, DAC3)

Resolution: 16 bits

Data Buffer: PC based memory

Output Voltage Range:

±10V

Output Current:

±1 mA max; sourcing more current may require a TR-2 power adapter option.

Offset Error:

±0.0045V maximum

Digital Feedthrough: <10 mV when updated

DAC Analog Glitch: <12 mV typical at major carry

Gain Error:

±0.01%

Update Rate: 1 MHz maximum, 19 hours minimum (no minimum with external clock); resolution: 20.83 ns.
Settling Time: 2

µs to rated accuracy

Clock Sources: 4 programmable

1.

Onboard D/A clock, independent of scanning input clock

2.

Onboard scanning input clock

3.

External D/A input clock, independent of external scanning input clock

4.

External scanning input clock

Digital I/O

One Digital I/O Channel, Typical

Channels: 24

Ports: 3 x 8-bit. Each port is programmable as input or output.

Input Scanning Modes: 2 programmable

1.

Asynchronous, under program control at any time relative to input scanning

2.

Synchronous with input scanning

Input Characteristics: 220

Ω series resistor, 20 pF to common

Logic Keeper Circuit: Holds the logic value to 0 or 1 when there is no external driver.

Input Protection:

±15 kV ESD clamp diodes parallel

Input Levels:

Low: 0 to 0.8V
High: +2.0V to +5.0V

Output Levels:

Low: < 0.8V
High: >2.0V

Output Characteristics: Output 1.0 mA per pin; ; sourcing more current may require

a TR-2 power adapter option.

Sampling Rate: 4 MHz maximum

Update Rate: 4 MHz maximum; 19 hours minimum (no minimum with external clock); resolution: 20.83 ns.