7-6 DaqBoard/3000USB Series Specifications

928691

Note: Specifications are subject to change without notice.

Triggering

Trigger Sources: 7, individually selectable for starting and stopping an acquisition. Stop acquisition can occur
on a different channel than start acquisition; stop acquisition can be triggered via modes 2, 4, 5, or 6 described
below.

1. Single-Channel Analog Hardware Trigger: Any analog input channel can be software programmed

as the analog trigger channel, including any of the analog expansion channels.

Input Signal Range: -10 to +10V max

Trigger Level: Programmable; 12-bit resolution

Hysteresis: Programmable; 12-bit resolution

Latency: 350 ns typical, 1.3 µs max

Accuracy: ±0.5% of reading, ±2 mV offset

Noise: 2 mV RMS

2. Single-Channel Analog Software Trigger: Any analog input channel, including any of the analog

expansion channels, can be selected as the software trigger channel. If the trigger channel involves a
calculation, such as temperature, then the driver automatically compensates for the delay required to
obtain the reading, resulting in a maximum latency of one scan period.

Input Signal Range: Anywhere within the range of the selected trigger channel

Trigger Level: Programmable; 16-bit resolution, including “window triggering”

Latency: One scan period max

3. Single-Channel Digital Trigger: A separate digital input is provided for digital triggering.

Input Signal Range: -15V to +15V

Trigger Level: TTL

Minimum Pulse Width: 50 ns high; 50 ns low

Latency: 100 ns typical, 1.1 µs max

4. Digital Pattern Triggering: 8 or 16-bit pattern triggering on any of the digital input ports.

Programmable for trigger on equal, above, below, or within/outside of a window. Individual bits can be
masked for “don’t care” condition.
Latency: One scan period max

5. Counter/Totalizer Triggering: Counter/totalizer inputs can trigger an acquisition. User can select to

trigger on a frequency or on total counts that are equal, above, below, or within/outside of a window.
Latency: One scan period max

6. Software Triggering: Trigger can be initiated under program control.

7. Multi-Channel Triggering: Up to 16 channels can be used to generate a trigger condition for any

combination of analog, digital, or counter inputs. Multiple channels can either be combined in a logical
“or” or “and” condition, with hysteresis programmable per channel. Maximum latency in this mode is
one scan period.