Labside

Labside adjustment of a temporary, chairside-modi-
fied emergence profile

– Screw the emergence profile modified by the operator

on a model analog.

– Take a silicone impression of the area of the emergence

profile to be adjusted.

– Remove the material applied by the operator.

Labside design of the emergence profile

Erase the area to be redesigned on the model with stone
or on the gingival mask.

Roughening, Telio Activator and SR Composiv

Apply SR Nexco.

With SR Nexco / Telio Lab LC (light-curing):

Blast the area to be supplemented (Al

2

O

3

, 100 µm, 1-2 bar / 15-29 psi). Then clean with steam and dry with oil-free

compressed air.

a) Use of Telio Activator or Telio Lab Cold Liquid

Condition with Telio Activator or Telio Lab Cold Liquid. For this purpose, distribute Telio Activator / Telio Lab Cold Liquid
evenly but generously on the conditioned surface and allow it to react for at least 2 min to a maximum of 4 min. Then,
apply SR Composiv according to the Instructions for Use, cure and subsequently layer the Telio Lab LC or SR Nexco
materials (see Telio CAD/Lab or SR Nexco Instructions for Use).