beautypg.com

Telio, Cad abutment solutions, Seating and aftercare – Ivoclar Vivadent Telio CAD A16 User Manual

Page 19

background image

19

Hybrid abutment crowns must be disinfected before being incorporated in the oral cavity. The local statutory provisions and
hygiene standards that apply to dental practices have to be observed. For disinfection, the hybrid abutment crown can be
immersed in a disinfectant suitable for PMMA materials. Dürr MD 520 with a reaction time of 5 min is a suitable impression
disinfectant.

Intraoral preparation

Please observe the following procedure to prepare for the permanent cementation of the implant-supported restoration:
– Remove the gingiva former or healing cap.
– Clean the implant lumen.
– Check the periimplant tissue (emergence profile).

Seating the hybrid abutment crown

Seating the hybrid abutment crown

– Insert the hybrid abutment crown intraorally into the implant.
– Manually screw in the matching implant screw.
– Tighten the implant screw with a torque wrench (observe the instructions of the manufacturer).
– Check the screw channel for contamination / moisture and clean or dry with an air syringe, if necessary.
– Insert a sterile cotton or foam pellet or teflon tape into the screw channel.
– Seal the screw channel with a composite or a light-curing temporary restorative (e.g. Telio CS Inlay / Onlay) if the wear

period is intended to be shorter than 6 weeks. For a longer wear period, a composite (e.g. Telio Add-On Flow, Tetric
EvoCeram, Tetric EvoFlow, Heliomolar) has to be used, at best after conditioning of the surface with Monobond Plus and
Heliobond.

– Polymerize with a curing light (e.g. Bluephase Style).
– Check the occlusion / articulation after polymerization and correct possible interfering spots with suitable fine-grain

grinding instruments.

– Polish to a high gloss with silicone polishers (e.g. OptraPol / Astropol).

Telio

®

CAD Abutment Solutions

Seating and Aftercare

Telio

®

CAD Abutment Solutions

– Seating and After

car

e