Operator's guide

In the five-stage process, each mass in turn is locked with a motorised

micrometer (stages 1, 2, 4), and the

N/S and E/W sensor bases are tilted

to their end stops (stages 3 and 5). At some point during each tilting

stage, the position of the relevant mass will flip to one or other side.

The

BUSY LED is lit during each stage, but goes out briefly between

stages, allowing you to follow the progress of the lock.

UNLOCK

This command unlocks the sensor masses and prepares the instrument
to begin operating.

If

UNLOCK is activated when the masses are already unlocked, the

processor will lock them and attempt to unlock again. This is useful if
you suspect that the locking procedure has failed.

During the

UNLOCK procedure, the instrument automatically

performs a round of centring for each component.