Engaging the hole lock – Guralp Systems CMG-3TB User Manual

Page 23

background image

Operator's guide

Caution: The hole lock may be using high-voltage mains (outlet)

power.

1. Connect the hole lock control unit to the

HOLELOCK POWER

connector of the breakout box, and to a mains power supply.

Alternatively, connect a 12 – 24 V DC power supply across the

input terminals of the hole lock control unit.

Do not connect

both DC and mains power at the same time.

The hole lock control unit supplied in regions with 220 V AC
mains power differs from that supplied for 110 V AC mains

power. You should ensure that you provide the correct voltage
to the hole lock control unit, otherwise damage may result to the

sensor.

2. Open the

MASS LOCK / UNLOCK – POWER ON / OFF flap on

the breakout box, and set the sensor power switch to the

off

position. The hole lock will

only function whilst the power is

off, to avoid injecting current transients from the mains power

supply into the sensor electronics.

3. If you are using a deep-borehole hole lock control unit, set the

dial to zero.

Engaging the hole lock

To extend the jaw of the hole lock:

1. Hold the switch on the hole lock control unit in the

EXTEND

JAW or + position. If you are using a deep-borehole control

November 2006

23