Operator's guide

Depending on the digitizer type, sine-wave, step and broadband noise

signal generators may be available. In this section, broadband noise
calibration will be used to determine the complete sensor response in

one action. Please refer to the digitizer's manual for information on
other calibration methods.

1. In Scream!'s main window, right-click on the digitizer's icon and

select Control.... Open the

Calibration pane.

2. Select the calibration channel corresponding to the instrument,

and choose Broadband Noise. Select the component you wish to

calibrate, together with a suitable duration and amplitude, and
click Inject now. A new data stream, ending Cn (n = 0 – 7) or
MB, should appear in Scream!'s main window containing the

returned calibration signal.

3. Open a Waveview window on the calibration signal and the

returned streams by selecting them and double-clicking. The

streams should display the calibration signal combined with the