CMG-3TB

The bbnoisecal script automatically performs appropriate

averaging to reduce the effects of aliasing and cultural noise.

Sensor response codes

Sensor

Sensor type code

Units

(V/A)

CMG-3TB, 30 s – 50 Hz response

CMG-3B_30S_50HZ

V

CMG-3TB, 100 s – 50 Hz response

CMG-3B_100S_50HZ

V

CMG-3TB, 120 s – 50 Hz response

CMG-3B_120S_50HZ

V

CMG-3TB, 360 s – 50 Hz response

CMG-3B_360S_50HZ

V

CMG-3TB, 360 s – 100 Hz response

CMG-3B_360S_100HZ

V

4.4 Calibration with a handheld control unit

If you prefer, you can inject your own calibration signals into the

system through a handheld control unit. The unit includes a switch
which activates the calibration relay in the seismometer, and 4 mm

banana sockets for an external signal source. As above, the equivalent
input velocity for a sinusoidal calibration signal is given by

v = V / 2 π f R K

where

V is the peak-to-peak voltage of the calibration signal, f is the

signal frequency,

R is the magnitude of the calibration resistor and K is

the feedback coil constant.

R and K are both given on the calibration

sheet supplied with the 3TB.

The calibration resistor is placed in series with the transducer.