CMG-3TB

4.2 Calibration methods

Velocity sensors such as the 3TB are not sensitive to constant DC
levels, either as a result of their design or because of an interposed

high-pass filter. Instead, three common calibration techniques are
used.

Injecting a step current allows the system response to be

determined in the time domain. The amplitude and phase
response can then be calculated using a Fourier transform.

Because the input signal has predominantly low-frequency
components, this method generally gives poor results. However,

it is simple enough to be performed daily.

Injecting a sinusoidal current of known amplitude and
frequency allows the system response to be determined at a spot

frequency. However, before the calibration measurement can be
made the system must be allowed to reach a steady state; for low

frequencies, this may take a long time. In addition, several
measurements must be made to determine the response over the

full frequency spectrum.

Injecting white noise into the calibration coil gives the response
of the whole system, which can be measured using a spectrum

analyser.

You can perform calibration either using a Güralp DM24 digitizer,
which can generate step and sinusoidal calibration signals, or by

feeding your own signals into the instrument through a handheld
control unit.

Before you can calibrate the instrument, its calibration relays need to

be activated by pulling low the

CAL ENABLE line on the instrument's

connector for the component you wish to calibrate. Once enabled, a

calibration signal provided across the

CAL SIGNAL and SIGNAL

GROUND lines will be routed through the feedback system. You can

then measure the signal's equivalent velocity on the sensor's output
lines. Güralp Handheld Control Units provide a switch for activating

the

CAL ENABLE line.

4.3 Calibration with Scream!

Güralp digitizers provide calibration signal generators to help you set

up your sensors. Calibration is most easily done through a PC running
Güralp's Scream! software.