beautypg.com

Guralp Systems CMG-3TB User Manual

Page 49

background image

Operator's guide

The N and E streams are the “North/South” and “East/West”

components of the downhole instrument. The X stream is the
North/South component of the reference instrument.

If you are using a separate digitizer, the reference instrument

will appear on the North/South component of the other digitizer,
instead of the X stream. To select this at the same time as the

other streams, make sure

Network is selected in the left-hand

pane of Scream!'s main window, to display all the streams from

your seismic network.

7. Double-click on one of the selected streams, or press ENTER. A

WaveView window will open.

8. Drag the streams across the window so that the reference stream

is at the top, the N stream in the middle, and the E stream at the

bottom.

9. Click the

Pause icon to stop the traces moving, and zoom in and

out until you can see a suitable data range. You should use a

period of at least an hour, and preferably longer.

10.Hold down the SHIFT key and drag across the WaveView

window with the left button until all three streams are selected

for the whole range.

Make sure there are no gaps in the data you select. In Scream!
4.3 and later, the selection will be shown with hatched lines if

there is a gap:

November 2006

49