beautypg.com

Poles and zeroes – Guralp Systems CMG-3TB User Manual

Page 56

background image

CMG-3TB

Poles and zeroes

Most users of seismometers find it convenient to consider the sensor as
a “black box”, which produces an output signal

V from a measured

input

x. So long as the relationship between V and x is known, the

details of the internal mechanics and electronics can be disregarded.

This relationship, given in terms of the Laplace variable

s, takes the

form

(

V / x ) (s) = G × A × H (s)

In this equation

G is the acceleration output sensitivity (gain constant) of the

instrument. This relates the actual output to the desired input
over the flat portion of the frequency response.

A is a constant which is evaluated so that A × H (s) is

dimensionless and has a value of 1 over the flat portion of the
frequency response. In practice, it is possible to design a system

transfer function with a very wide-range flat frequency
response.

The normalising constant

A is calculated at a normalising

frequency value

fm = 1 Hz, with s = j fm, where j = √–1.

H (s) is the transfer function of the sensor, which can be

expressed in factored form:

In this equation

z

n

are the roots of the numerator polynomial,

giving the zeros of the transfer function, and

p

m

are the roots of

the denominator polynomial giving the poles of the transfer
function.

In the calibration pack,

G is the sensitivity given for each component

on the first page, whilst the roots

z

n

and

p

m

, together with the

normalising factor

A, are given in the Poles and Zeros table. The poles

and zeros given are measured directly at Güralp Systems' factory using
a spectrum analyser. Transfer functions for the vertical and horizontal

sensors may be provided separately.

56

Issue C

This manual is related to the following products: