2 the control system, Lock – Guralp Systems CMG-3TB User Manual

Page 66

background image

CMG-3TB

5.2 The control system

The internal operations of the 3TB are supervised by a control
microprocessor, which drives the mass clamping and centring

adjustment motors. It responds to commands sent on three input lines
by grounding for 0.2 – 7 seconds.

The signals you can send to the microprocessor are termed

LOCK,

UNLOCK, and CENTRE. Each command acts on the vertical, N/S and

E/W masses in turn. The microprocessor prevents the system from

attempting incompatible actions (

e.g. centring when the masses are

clamped.)

While a command is taking place, the

BUSY LED will flash; you can

use this for diagnostic purposes. See the description of each command
for full details.

When no command is active,

i.e. all three lines are high, the control

microprocessor goes into a power-saving mode. In routine operation,
the lines are controlled from the breakout box, Handheld Control Unit

or digitizer. If you send control signals to the 3TB manually, you must
ensure that the lines are pulled high after sending the signal, or the

equipment may be damaged. A “biased-OFF” type switch can be used
for this purpose.

LOCK

This command unlocks the masses and clamps the horizontal sensors
by tilting them up to their end stops.

If

LOCK is activated when the masses are already locked, the processor

will unlock them and attempt to lock again. This is useful if you
suspect that the locking procedure has failed.

In detail, the process acts as shown in the following graph. The top

three streams are the

mass position outputs of each component (Z, N/S

and E/W, respectively), whilst the bottom one represents the state of

the

BUSY LED (up = on).

66

Issue C

This manual is related to the following products: