beautypg.com

Frequency response curves, Obtaining copies of the calibration pack – Guralp Systems CMG-3TB User Manual

Page 57

background image

Operator's guide

Frequency response curves

The frequency response of each component of the 3TB is described in
the normalised amplitude and phase plots provided. The response is

measured at low and high frequencies in two separate experiments.
Each plot marks the low-frequency and high-frequency cutoff values

(also known as –3 dB or half-power points).

If you want to repeat the calibration to obtain more precise values at a

frequency of interest, or to check that a sensor is still functioning
correctly, you can inject calibration signals into the system using a

Güralp digitizer or your own signal generator, and record the
instrument's response.

Obtaining copies of the calibration pack

Our servers keep copies of all calibration data that we send out. In the
event that the calibration information becomes separated from the

instrument, you can obtain all the information using our free e-mail
service. Simply e-mail

[email protected]

with the serial number

of the instrument in the subject line,

e.g.

From: [email protected]

To: [email protected]

Subject: T3A15

The server will reply with the calibration documentation in Word

format. The body of your e-mail will be ignored.

November 2006

57

This manual is related to the following products: