Bronkhorst EL-FLOW Prestige User Manual
Page 34
![background image](https://www.manualsdir.com/files/776608/content/doc034.png)
Bronkhorst High-Tech B.V.
EL-FLOW Prestige
9.17.084
34
Reset
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char
R
0…5
114
115/8
0x0E68/3689
This parameter is used to reset program, counter or alarms.
Value 0: no reset
Value 1: Reset counter
Value 2: Reset alarm
Value 3: Reset counter
Value 4: Reset counter and counter off
Value 5: Reset firmware program (soft reset)
Make sure the ‘
114
Reset’ value is accepted by sending a 0 first. The ‘
114
Reset’ parameter may be disabled by the
‘
156
Reset Alarm Enable’ or ‘
157
Reset Counter Enable’ parameters.
Wink
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char [27]
W
0…9
1
0/0
0x0000/1
Any text string value between '1' and '9' lets the instrument wink (normal wink) (red and green turn by turn) for that number of
seconds for identifying its position. Default value = '0'.
Control Mode
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned int
RW
0…255
12
1/4
0x0024/37
The ‘
12
Control Mode’ is used to select different controlling modes of the instrument and determines from which sources controller
setpoints are accepted. The following control modes are available:
Value
Mode
Instrument action
Setpoint source
Slave factor
0
BUS/RS232
Normal operation, controlling
Bus or RS232
1
Analog Input
Normal operation, controlling
Analog input
2
FLOW-BUS Slave
Controlling as slave of other instrument
on bus
‘FLOW-BUS master output’ x
‘
139
Slave Factor’ / 100%
‘
139
Slave Factor’
3
Valve Close
Close valve
4
Controller Idle
Stand-by on bus/RS232, controlling is
stopped; Valve Output freezes in current
position
7
Setpoint 100%
Controlling at setpoint 100%
Fixed 100%
8
Valve Fully Open
Valve fully opened
9
Calibration Mode
Calibration mode enabled
(factory only)
10
Analog Slave
Controlling as slave of other instrument
on analog input
‘
11
Analog Input’ x ‘
139
Slave
Factor’ /100%
‘
139
Slave Factor’
12
Setpoint 0%
Controlling at setpoint 0%
Fixed 0%
13
FLOW-BUS
Analog Slave
Controlling as slave of other instrument
on bus, slave factor is set with signal on
analog input
‘FLOW-BUS master output’ x
‘
11
Analog Input’ x ‘
139
Slave
Factor’ / 100%
‘
11
Analog Input’
18
RS232
Normal operation, controlling
RS232