beautypg.com

Software (dde applications), Baud rate setup, 3 software (dde applications) – Bronkhorst EL-FLOW Prestige User Manual

Page 19: 4 baud rate setup

background image

Bronkhorst High-Tech B.V.

EL-FLOW Prestige

19

9.17.084

3.4.3

Software (DDE applications)

Examples of free Bronkhorst DDE client applications: FlowDDE, FlowPlot and FlowView. Other software programs supporting DDE

are for example MS-Office, LabVIEW, InTouch and Wizcon.

Bronkhorst® software applications 'FlowView' (left) and 'FlowPlot' (right)

i

www

FlowDDE and other Bronkhorst applications are available on the support CD or can be downloaded from the

Bronkhorst internet site: http://www.bronkhorst.com/en/products/accessories_and_software/

bronkhorst_flowware/

3.4.4

Baud rate setup

EL-FLOW Prestige instruments support the following baud rates for RS232 communication. The factory selected baud rate is

indicated on the instrument back-side label. Refer to section 4.3.2 for changing the baud rate settings for the instrument. The

default baud rate for RS232 communication is 38400 Baud.

Mode:

Digital

Interface/medium:

RS232

Protocol:

FLOW-BUS (Propar)

Baud rate:

9600

16200

38400

57600

115200

230400

460800

Node address:

3

Parity:

None

RS232 communication options

Changing RS232 settings of the 9-pin sub-D side connector interface

Refer to section 4.3.2 for changing the baud rate settings for the instrument.

Make sure that the instrument’s baud rate corresponds with the baud rate of the application the instrument is

communicating with, otherwise no communication can be established.

For RS232 communication, the maximum cable length is 10 m for baud rates up to 38400 Baud. For higher baud

rates, use cable lengths of maximum 3 m.

i

www

For more information regarding communication through an RS232 interface, see document 9.17.027: RS232

interface with FLOW-BUS for digital instruments. http://www.bronkhorst.com/files/downloads/

manuals_english/917027manual_rs232_interface.pdf