beautypg.com

Micro switch functions, 8 micro switch functions – Bronkhorst EL-FLOW Prestige User Manual

Page 26

background image

Bronkhorst High-Tech B.V.

EL-FLOW Prestige

9.17.084

26

3.8

Micro switch functions

By means of manual operation of the micro push-button switch some important actions for the instrument can be selected or

started. These options are available in both analog and digital operation mode. These functions are:

Reset alarm
Reset instrument (firmware program reset)
Auto-zero
Restore factory settings (in case of accidentally changing of the settings)
Activate 'Configuration Mode' (for changing communication settings via RS232)

Using digital RS485 operation via the 9-pin sub-D side connector it is also possible to read/set:

Bus-address (node-address) (only required for RS485)

The micro switch on top of the EL-FLOW Prestige can be operated with a thin, metal or hard plastic pin, for example

the end of a paperclip.

When the micro switch is pressed, both LEDs will start indicating different patterns in a loop. The switch has to be pressed down

until the two LEDs are indicating the right pattern. When the switch is released, the selected action is started. The tables below

describe the micro switch functions that can be started in normal operation mode and during power-up:

LEDs

Time

pushed

Indication

green ‘Mode’ LED

red ‘Error’ LED

Off

Off

0…1 sec.

No action

Pressing a switch briefly by accident will not start any unwanted

reaction of the instrument.

Off

Off

1…4 sec.

In case of min/max alarm or counter batch reached: Reset alarm

(only if reset by micro switch has been enabled).

For FLOW-BUS only: if the node address is occupied, this function

will install a free node-address on FLOW-BUS.

Off

On (red)

4…8 sec.

Reset instrument

Instrument program will be restarted and all warning and error

messages will be cleared. During start-up the instrument will

perform a (new) self-test.

On (green)

Off

8…12 sec.

Auto-zero

Instrument will be re-adjusted for measurement of zero-flow, see

section 2.9.

On (green)

On (red)

12…16

sec.

Prepare instrument for FLASH mode for firmware update.

Instrument shuts down and both LEDs turn off. At next power-up the

instrument will be active again.

LED indications using micro switch at normal operation mode of an instrument

LEDs

Time

pushed

Indication

green ‘Mode’ LED

red ‘Error’ LED

Off

Off

0…4 sec.

No action

Pressing a switch briefly by accident will not start any unwanted

reaction of the instrument.

Off

Normal flash

0.2 sec on,

0.2 sec off

4…8 sec.

Restore factory settings

All parameter settings (except field bus/communication settings) will

be restored to the original factory settings.

Normal flash

0.2 sec on, 0.2 sec off

On (red)

8…12 sec.

For FLOW-BUS only: install a free node-address on FLOW-BUS.

Normal flash

0.2 sec on, 0.2 sec off

Normal flash

0.2 sec on,

0.2 sec off

12…16

sec.

Activate ‘Configuration Mode’

The baud rate and bus type for the 9-pin sub-D side connector are

set to 38k4 and RS232 FLOW-BUS (Propar). The ‘Configuration Mode’

is recognized by the green LED blinking 2 sec on, 0.1 sec off. The

‘Configuration Mode’ is deactivated only after applying this micro

switch action again.

LED indications using micro switch at power-up situation of an instrument