distance between the angle and the instrument is recommended.

Do not mount pressure regulators directly on the inlet of gas flow meters/controllers, but allow at least 25 pipe

diameters of piping.

2.4

Instrument mounting

Install the EL-FLOW Prestige meter/controller in the line, in accordance with

the direction of the FLOW arrow. The arrow for flow direction is indicated on

the body of the instrument. Tighten the fittings according to the instructions of

the supplier of the fittings. The use of Swagelok RS-type stainless steel adapters

is recommended.

For EL-FLOW Prestige instruments the upright position is preferred. When mounting an instrument in up- or

downward position, zeroing of the instrument is advised (see section 2.9), especially at high pressures (> 10 bar).

Avoid installation in close proximity of mechanic vibration and/or heat sources.

At the bottom side of the instrument a number of mounting holes are present, for stable fixation of the instrument.

Refer to the dimensional drawing for the instrument for the exact position of the mounting holes.

i

www

Dimensional drawings can be found at: http://www.bronkhorst.com/en/downloads

2.5

Leak check

Check the system for leaks before applying (fluid) pressure. Especially if toxic, explosive or other dangerous fluids are

used!

2.6

Electrical connection

Electrical connection must be made with standard cables or according to the applicable

hook-up diagrams. The factory installed 9-pin sub-D settings are indicated on the

instrument back-side label. Make sure that the power supply is suitable for the power

ratings as indicated on the instrument label and that double or reinforced insulation is

used for the power supply.

i

www

Applicable hook-up diagrams for EL-FLOW Prestige can be found at:

http://www.bronkhorst.com/en/downloads

EL-FLOW Prestige instruments are powered with +15…+24 Vdc. Several

hook-up examples and standard cables are found in chapter 3.

The instruments contain electronic components that are susceptible to damage by electrostatic discharge. Proper

handling procedures must be taken during installation, removing and connecting the electronics.

The instruments described in this manual carry the CE-mark and are compliant with the EMC requirements. However

compliance with the EMC requirements is not possible without the use of proper cables and connector/gland

assemblies. Bronkhorst High-Tech B.V. recommends the use of their standard cables. These cables have the right

connectors and if loose ends are used, these will be marked to prevent wrong connection. When using other cables,

cable wire diameters should be sufficient to carry the supply current and voltage losses must be kept as low as

possible. When in doubt: contact your distributor.

When connecting the system to other devices (e.g. PLC), be sure that the integrity of the shielding is not affected. Do

not use unshielded wire terminals.