Advanced parameters and properties, Advanced measurement and control parameters, Special instrument parameters – Bronkhorst EL-FLOW Prestige User Manual
Page 33: 2 advanced parameters and properties, 1 advanced measurement and control parameters, 2 special instrument parameters
Bronkhorst High-Tech B.V.
EL-FLOW Prestige
33
9.17.084
4.2
Advanced parameters and properties
4.2.1
Advanced measurement and control parameters
Measured Value (Fmeasure)
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Float
R
-3.4E+38…
3.4E+38
205
33/0
0xA100…0xA101/ 41217…41218
Floating point variable of the ‘
8
Measured Value’. The ‘
205
Fmeasure’ parameter shows the measured value in the capacity and
capacity unit for which the instrument has been set. The ‘
205
Fmeasure’ parameter is dependent of ‘
129
Capacity Unit’ and ‘
22
Sensor
Type’.
Setpoint (Fsetpoint)
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Float
RW
0…3.4E+38
206
33/1
0xA119…0xA11A/ 41241…41242
Floating point variable of the ‘
9
Setpoint’. The ‘
206
Fsetpoint’ parameter shows the setpoint in the capacity and capacity unit for
which the instrument has been set. The ‘
206
Fsetpoint’ parameter is dependent of ‘
129
Capacity Unit’ and ‘
22
Sensor Type’.
Valve Output
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned long
RW
0…
16777215
55
114/1
0x001F/32
This parameter is the digital steering signal for driving the control valve, where 0…16777215 corresponds with 0…100%.
Analog Input
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned int
R
0…65535
11
1/3
0x0023/36
Depending on the analog mode, 0…5 Vdc, 0…10 Vdc, 0…20 mA or 4…20 mA is converted to a number in the range 0…32000. The
digitized ‘
11
Analog Input’ is in the same range as ‘
8
Measured Value’ (0…32000 corresponds to 0…100%). This parameter can be used
as setpoint or slave factor when the instrument is used as ‘analog slave’ or ‘FLOW-BUS analog slave’, see ‘
12
Control Mode’.
4.2.2
Special instrument parameters
All parameters described in this section have influence on the behavior of the EL-FLOW Prestige. Please be aware
that wrong settings can disorder the output. To avoid unintentional changes, some parameters are locked (shown by
the symbol). To unlock parameters set parameter ‘
7
Init Reset’ to ‘Unlocked’.
Init Reset
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char
RW
82/64
7
0/10
0x000A/11
The ‘
7
Init Reset’ parameter is used to unlock secured parameters for writing (see the symbol). This parameter can be set to the
following values:
Value 82: Locked mode, secured parameters are read-only
Value 64: Unlocked mode, secured parameters are writeable and readable
This parameter is always set to ‘Locked’ (value 82) at power-up.