Isol setup and operating instructions, Isol calibration – AMETEK XHR 1000 Watt Series User Manual

Page 73

background image

Remote Operation

Using the Isolated Programming (ISOL) Interface Option

73

ISOL Setup

and

Operating

Instructions

1. Ensure that the rear panel SW1 switch is set to its default configuration as in

Figure 1.4 , on p. 19.

2. Connect the 0-5 V programming source to the J2 connector on the power

supply’s rear panel as described in

Section 4. Remote Operation

.

3. Once the connections are made, use the power supply with the ISOL interface

installed as you would a power supply with a 0-5 Vdc programming source. The
only differences are that the programming and monitoring sources are now
isolated and that resistive and 0-10 V programming as well as 0-10 V monitoring
are unavailable. All front panel functions are as usual.

ISOL

Calibration

To calibrate a unit which has the ISOL interface installed, have the service technician
follow the standard procedures in Section 5. Calibration.

!

CAUTION
In order to maintain isolation, do not connect either J2 connector terminal 3
(control ground) or the power supply return to J2 terminal 6 (program/monitor
signal return).