Remote Operation
Using the Isolated Programming (ISOL) Interface Option

72

Operating Manual for XHR Series Power Supply

Figure 4.4 0-10 Vdc Output Voltage and Current Monitoring

Using the Isolated Programming (ISOL) Interface Option

With the 4-channel Isolated (ISOL) Programming Interface installed, you can
control and monitor the power supply’s output voltage and current limit using 0-5 V
analog signals that are not referenced to the power supply’s internal ground. The
isolation amplifiers used are rated to 1000 Vdc. With ISOL installed, the power
supply will no longer support resistive programming, or 0-10 V programming or
readback. You can still control any non-programmed function (either voltage or
current limit) at the power supply’s front panel.

ISOL

Specifications

Nominal Gain

1 V/V typical

Gain vs. Temperature

±10 ppm/°C typical

Non Linearity

±0.016% FSR typical, ±0.020%
FSR maximum

Offset Voltage

±20 mV typical, ±50 mV
maximum

Input vs. Temperature

±200 mV/°C typical

Input Voltage Range

0-5 V

Input Impedance

200 k typical

Readback Voltage Range

0-5 V

Readback Current Drive

±15 mA typical, ±5 mA minimum

Readback Load Capacitance

0.1 mF typical

SW1-5 and SW1-6

OFF (OPEN)