Remote Monitoring of Output Voltage and Current

71

Remote Monitoring of Output Voltage and Current

Readback

Signals

The J2 connector on the rear panel provides access to calibrated readback signals for
remote monitoring of the output voltage and current. Use rear panel switches SW1-5
and SW1-6 to select either a 0-5 Vdc or a 0-10 Vdc range for the output. The
readback signal represents 0 to 100% of the power supply’s output.

See

Table 4.2

for the required J2 connections and switch settings for remote

monitoring of readback signals with 0-5 Vdc or 0-10 Vdc outputs. Use shielded pair
wiring (20 to 24 AWG) and ground the shield to J2 connector terminal 1 or to the
chassis. See

Figure 4.3

and

Figure 4.4

for sample configurations using remote

monitoring.

Table 4.2 Power Supply Settings for Remote Monitoring of Readback Signals

Note:Check the readback accuracy any time you reset the SW1 switch. Access offset
and range adjustment for the readback signals through the cover of the unit.
See “Calibrating for Readback Accuracy” on page 80.

Figure 4.3 0-5 Vdc Output Voltage and Current Monitoring

Readback Signal

J2 Connection
Signal (+)

J2 Connection
Return (-)

Switch SW1 Setting

Output Voltage
(0-5 Vdc)

J2-9

J2-6

SW1-5 ON (CLOSED)

Output Voltage
(0-10 Vdc)

J2-9

J2-6

SW1-5 OFF (OPEN)

Output Current
(0-5 Vdc)

J2-10

J2-6

SW1-6 ON (CLOSED)

Output Current
(0-10 Vdc)

J2-10

J2-6

SW1-6 OFF (OPEN)

SW1-5 and SW1-6:

ON (CLOSED)