Maintenance, Cleaning – AMETEK XEL Series 60V and below User Manual

Page 44

background image


Step Settings Display

Adjust

for

Load

Detail

1.

O/P off

Cal count

0V ± 2mV

DVM

O/P off volts

2.

0V

Cal count

0V ± 2mV

DVM

V DAC zero

3.

Max V, 100mA

Cal count

15/30/60V ± 2mV

DVM

V DAC scale

4.

Max V, 100mA

O/P volts

Display = DVM

DVM

V metering scale

5.

I pot at 0

I pot

Display = 0

-

I pot ADC zero

6.

2V, 0mA

Cal count

0mA ± 0.2mA

milli-ammeter

High I DAC zero

7.

2V, 0mA

O/P amps Display = milli-ammeter milli-ammeter

High I metering zero

Change

load

8.

2V, max I

Cal count

1.5/3A ± 1mA

Ammeter

High I DAC scale

9.

2V, max I

O/P amps

Display = ammeter

Ammeter

High I metering scale

10.

2V, 0mA

Cal count

0mA ± 0.02mA

milli-ammeter

Low I DAC zero

11.

2V, 0mA

O/P amps Display = milli-ammeter milli-ammeter

Low I metering zero

12.

2V, 500mA

Cal count

500mA ± 0.1mA

ammeter

Low I DAC scale

13.

2V, 500mA

O/P amps

Display = ammeter

ammeter

Low I metering scale

14.

O/P off

‘End’

-

-

Stores and exits

Maintenance

The Manufacturers or their agents overseas will provide a repair service for any unit developing a
fault. Where owners wish to undertake their own maintenance work, this should only be done by
skilled personnel in conjunction with the service manual which may be purchased directly from
the Manufacturers or their agents overseas.

Cleaning

If the instrument requires cleaning use a cloth that is only lightly dampened with water or a mild
detergent.

WARNING! TO AVOID ELECTRIC SHOCK, OR DAMAGE TO THE INSTRUMENT, NEVER
ALLOW WATER TO GET INSIDE THE CASE. TO AVOID DAMAGE TO THE CASE NEVER
CLEAN WITH SOLVENTS.